Go back to datasheets

PEEK-OPTIMA™ Wear Performance (MOTIS) (Granules)

High performance biocompatible thermoplastic material, 30% milled pitch carbon fibre reinforced PolyEtherEtherKetone (PEEK), semi crystalline. Granules for injection moulding and extrusion, standard flow, for use in long term human implantation. Colour black.

Typical Application Areas
Milled pitch carbon fiber composite, specifically designed for tribological applications particularly against a hard counterface such as metal or ceramic whilst maintaining lightweight and radiolucent properties.

Suitable for use in long-term implantable medical devices. Excellent sterilisation resistance. As PEEK is hygroscopic, drying before use is recommended. Further information is available upon request.

Material Properties
Physical

Nominal Value

Unit

Test Method

Density (23°C)

1.42

g/cm³

ISO 1183

Molding Shrinkage

ASTM D955

Flow

0.30

%

Across Flow

0.70

%

Water Absorption (Equilibrium, 23°C, 50% RH)

0.50

%

ISO 62

Crystallinity DSC

30.0

%

Internal Method
Mechanical

Nominal Value

Unit

Test Method

Tensile Stress

ISO 527-2

Yield, 23°C 1

152

MPa

Yield, 23°C 2

147

MPa

Yield, 23°C 3

149

MPa

Yield, 23°C 4

145

MPa

Tensile Strain 4 (Break, 23°C)

2.2

%

ISO 527-2

Flexural Modulus

23°C 1

11400

MPa

ISO 178

23°C 3

11300

MPa

ISO 178

23°C 4

11600

MPa

ISO 178

Flexural Stress

Yield, 23°C 1

231

MPa

ISO 178

Yield, 23°C 2

228

MPa

ISO 178

Yield, 23°C 4

227

MPa

ISO 178

Compressive Modulus 4 (23°C)

12000

MPa

ISO 604

Compressive Stress 4 (23°C)

200

MPa

ISO 604

Shear Modulus 4 (23°C)

2200

MPa

ISO 15310

Shear Strength 4 (23°C)

95.0

MPa

ASTM D732

Poisson's Ratio 4 (23°C)

0.41

ASTM D638
Impact

Nominal Value

Unit

Test Method

Notched Izod Impact Strength

ISO 180

23°C 1

6.4

kJ/m²

23°C 2

6.6

kJ/m²

23°C 3

6.2

kJ/m²

23°C 4

6.7

kJ/m²

Hardness

Nominal Value

Unit

Test Method

Rockwell Hardness (M-Scale, 23°C)

104

ISO 2039-2
Thermal

Nominal Value

Unit

Test Method

Glass Transition Temperature (Onset)

140

°C

ISO 11357-2

Melting Temperature

339

°C

ISO 11357-3

CLTE - Flow

ASTM D696

50 to 120°C

8.0E-7

cm/cm/°C

170 to 220°C

1.5E-6

cm/cm/°C

Recrystallization Temperature (Peak)

284

°C

ISO 11357-3
Electrical

Nominal Value

Unit

Test Method

Volume Resistivity

ASTM D257

-- 5

4.1E+9

ohms·cm

-- 6

2.1E+9

ohms·cm

Fill Analysis

Nominal Value

Unit

Test Method

Melt Viscosity 1000 s^-1 (400°C)

730

Pa·s

Internal Method
Typical Processing Information
Injection

Nominal Value

Unit

Drying Temperature

120 to 150

°C

Drying Time

3.0 to 5.0

hr

Suggested Max Moisture

0.020

%

Hopper Temperature

< 100

°C

Rear Temperature

375

°C

Middle Temperature

380 to 385

°C

Front Temperature

390

°C

Nozzle Temperature

395

°C

Mould Temperature

180 to 210

°C

Injection Notes

Drying Temperature / Time: 150°C / 3h or 120°C / 5h (residual moisture <0.02%)
Runner: Die / Nozzle >3 mm, Manifold >3.5 mm
Gate: >2 mm or 0.5 x part thickness

Important Notes:
1) Processing conditions quoted in our datasheets are typical of those used in our processing laboratories

  • Data for mould shrinkage should be used for material comparison. Actual mould shrinkage values are highly dependent on part geometry, mould configuration, and processing conditions.
  • Mould shrinkage differs for along flow and across flow directions. “Along flow” direction is taken as the direction the molten material is travelling when it exits the gate and enters the mould.
  • Mould shrinkage is expressed as a percent change in dimension of a specimen in relation to mould dimensions.

2) Data are generated in accordance with prevailing national, international and internal standards, and should be used for material comparison. Actual property values are highly dependent on part geometry, mould configuration and processing conditions. Properties may also differ for along flow and across flow directions.

Detailed data available on our website www.invibio.com or upon request.

Notes
11-Cycle EtO
23-Cycles Steam
375 kGy Gamma
4As Moulded
50.5 V
61.0 V
PEEK-OPTIMA Wear Performance MOTIS Granules