Go back to datasheets

PEEK-OPTIMA™ HA Enhanced LT120HA FIL 1.75mm

High performance thermoplastic material, PolyEtherEtherKetone (PEEK) as semi crystalline filament with hydroxyapatite content for Additive Manufacture by filament fusion and other melt extrusion 3D printing processes. Colour natural/beige.

Typical Application Areas
Filament Fusion Additive Manufacture of printed parts. Suitable for use in the 3D printing of long-term implantable medical devices. Excellent sterilisation resistance. Product supplied vacuum packed and dry when produced. As PEEK is hygroscopic, drying before use is recommended. This product is based on PEEK-OPTIMA™ LT120HA.

Material Properties
Physical

Nominal Value

Unit

Test Method

Density

1.47

g/cm³

ISO 1183

Linear Density

3.53

g/m

Internal Method

Filament Diameter 1

1.75

mm

Thermal

Nominal Value

Unit

Test Method

Glass Transition Temperature (Onset)

148

°C

ISO 11357-2

Melting Temperature

338

°C

ISO 11357-3

Recrystallization Temperature (Peak)

286

°C

ISO 11357-3
Fill Analysis

Nominal Value

Unit

Test Method

Melt Viscosity 1000 s^-1 (400°C)

566

Pa·s

Internal Method

Melt Stability 1000 s^-1, 1 hr (400°C)

2.0

%

Internal Method

Shear Viscosity 100 s^-1 (400°C)

1167

Pa·s

Internal Method

Shear Viscosity 1000 sec^-1 (400°C)

440

Pa·s

Internal Method

Shear Viscosity 10000 sec^-1 (400°C)

136

Pa·s

Internal Method
Additional Information
  • Nominal Weight: 0.5 kg
  • Nominal Length: 142 m


Packaging

  • Spool Dimensions: 200 mm diameter
  • Spool Dimensions: 68 mm width
  • Spool Material: Heat-resistant Polycarbonate
Typical Processing Information
Extrusion

Nominal Value

Unit

Drying Temperature

100

°C

Drying Time

8.0

hr

Suggested Max Moisture

0.020

%

Melt Temperature

340 to 450

°C

Extrusion Notes

Drying Temperature / Time: 100 °C / 8 h (residual moisture <0.02%)
Chamber / Build-Space Temperature: >150°C
Gate: >150°C

Important Notes
1) Processing conditions quoted in our datasheets are typical of those used in our processing laboratories.
2) Data are generated in accordance with prevailing national, international and internal standards, and should be used for material comparison. Actual property values are highly dependent on part geometry and processing conditions.

Detailed data available on our website www.invibio.com or upon request.

Notes
13 axis laser micrometre
PEEK OPTIMA Ha Enhanced Filament